| Liner Features
CERAMIC INSERT DESIGN
ALUMINA CERAMIC |
Alumina (Al2O3) was selected as the material for
the liner based on the history of its clinical use and its excellent
wear characteristics when paired with an alumina femoral head. It has
been used by the orthopaedic device industry since 1969 and has
exhibited excellent biocompatibility.
11,12
Most importantly, alumina on alumina articulation couples have demonstrated significantly less wear
than THAs involving a polyethylene socket.
13
LOCKING MECHANISM |
The ceramic LINEAGE® insert and shell are
assembled at the time of surgery by impacting the insert into the shell.
A self-locating, eighteen degree taper provides a secure method of
assembly between the shell and insert. The strength of this shell/insert
construct is confirmed by the experimental results presented later in
this monograph. |